Reported Aug 27, 2025 · Ongoing

Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep…

Boston Scientific Neuromodulation Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Boston Scientific Neuromodulation Corporation · 25155 Rye Canyon Loop · Valencia, CA · 91355-5004

Is this the product you have?

  • Burr Hole Cover Kit
  • REFM365DB4600C0; Burr Hole Cover Spares Kit
  • REFM365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
Lot codes and dates
  • All lots. Surgical
  • Implant Manual version REF/UDI-DI
    • M365DB4600C0/08714729820802
    • M365DB4605C0/08714729820819
Amount recalled
24,583

Where it was sold

Why it was recalled

Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Boston Scientific Neuromodulation Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 3:23 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.