Reported Aug 10, 2016 · Terminated
Merge HEMO software.
Merge Healthcare, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Merge Healthcare, Inc. · 900 Walnut Ridge Dr · Hartland, WI · 53029-8347
Is this the product you have?
- Merge HEMO software.
- Lot codes and dates
Software versions 8.109.109.20
- Amount recalled
- 37
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
The International Normalized Ratio (INR) value displayed in the study report does not match the value that is imported on the pre-procedure labs screen from HL7
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Merge Healthcare, Inc. recalls