Reported Jun 17, 2020 · Terminated

RayStation 6, UDI # 07350002010013 - Product Usage: For radiation therapy and medical oncology which can also be used…

Raysearch Laboratories Ab

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: RAYSEARCH LABORATORIES AB · Sveavagen 9 · Stockholm, N/A · N/A

Is this the product you have?

    • RayStation 6
    • UDI # 07350002010013 -
  • Product UsageFor radiation therapy and medical oncology which can also be used to administer treatments
Lot codes and dates
  • UDI # 07350002010013
  • Build number6.0.0.24
Amount recalled
N/A

This recall covers 15 products

The agency filed one notice per product. You are reading one of them; the other 14 are below.

2 more are in the archive.

Where it was sold

Why it was recalled

Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Raysearch Laboratories Ab recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 6, 2021 · Data as of Sep 19, 4:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.