Reported Jul 11, 2018 · Terminated

microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA

FHC, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: FHC, Inc. · 1201 Main St · Bowdoin, ME · 04287-7302

Is this the product you have?

  • microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
Lot codes and dates
  • UDI00873263002162
Amount recalled
2952 total

This recall covers 15 products

The agency filed one notice per product. You are reading one of them; the other 14 are below.

2 more are in the archive.

Where it was sold

Why it was recalled

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new FHC, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 30, 2019 · Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.