Reported Aug 17, 2016 · Terminated

VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog…

Ortho-Clinical Diagnostics

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ORTHO-CLINICAL DIAGNOSTICS · FELINDRE MEADOWS · Bridgend · N/A

Is this the product you have?

  • VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog Number/Product Code 855 2630; IVD -- Ortho-Clinical Diagnostics, Pencoed, Bridgend CF35 5PZ, UK --- The VITROS Estradiol test is performed using the VITROS Estradiol Reagent Pack and Calibrators on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems.
Lot codes and dates
  • Lot numbers1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry date 10-Oct-16), 1510 (expiry date 01-Nov-16), 1528 (expiry date 01-Nov-16) 1538 (expiry date 14-Mar-17), 1548 (expiry date 09-Mar-17), 1558 (expiry date 09-Mar-17) and 1568 (expiry date 29-Mar-17)
Amount recalled
Domestic: 1620; Foreign: 7597

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug Fulvestrant and theVITROS Immunodiagnostic Estradiol Reagent assay. The firm conducted an internal investigation and was able to confirm biased positive (ranging from 1303-4046%) on Estradiol results, obtained from postmenopausal females taking Fulvestrant.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 8, 2018 · Data as of Sep 19, 6:01 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.