Reported Aug 17, 2016 · Terminated

Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product…

Hill-Rom, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Hill-Rom, Inc. · 1069 State Road 46 E · Batesville, IN · 47006-7520

Is this the product you have?

  • Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered
  • Adjustable Hospital Bed Product UsageVolker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims
Lot codes and dates
  • All beds (serial numbers) in the 2080, 2082, S 280 and S 282 series are affected by this field action except for those without siderails
  • The models are as follows
    • 2080
    • 2082
    • 2080 C
    • 2080 E
    • 2080 FS
    • 2080 K
    • 2080 LP
    • 2080 MA
    • 2080 MB
    • 2080 S
    • 2080 SL
    • 2080 SONDER
    • 2082 SONDER
    • S280
    • S280 SONDER
    • S282
    • S282 SONDER
Amount recalled
24262

Where it was sold

  • Nationwide

Why it was recalled

The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Hill-Rom, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 15, 2017 · Data as of Sep 19, 4:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.