Reported Sep 4, 2019 · Terminated
AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale…
Ad-Tech Medical Instrument Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ad-Tech Medical Instrument Corporation · 400 W Oakview Pkwy · Oak Creek, WI · 53154-7213
Is this the product you have?
- AD-TECH Spencer
- Probe Depth Electrode Product UsageThe Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping
- Lot codes and dates
- Catalog Numbers SD08R-SP05X-000 SD06R-SP05X-000 LOT NUMBERS 127219 208140699 127357 208140699
- Amount recalled
- 18
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
- IL
Why it was recalled
An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ad-Tech Medical Instrument Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.