Reported Jun 17, 2026 · Ongoing

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr · Newark, DE · 19702-2466

Is this the product you have?

  • Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Lot codes and dates
  • Manufacturing Lot/Serial Numbers
    • FA6254
    • FA6260
    • FA6286
  • UDIs
    • (01)00842768035425(10)FA6254(17)20260911
    • (01)00842768035425(10)FA6260(17)20260917
    • (01)00842768035425(10)FA6286(17)20261013
Amount recalled
2878 units

Where it was sold

  • Nationwide

Why it was recalled

A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:16 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.