Reported Sep 8, 2021 · Terminated
TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical…
DePuy Orthopaedics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: DePuy Orthopaedics, Inc. · 700 Orthopaedic Dr · Warsaw, IN · 46582-3994
Is this the product you have?
- TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking
- of bone before cutting Part Number420916
- Lot codes and dates
- Lot number/ GTIN
28226281352761227544274312740726970
- OUS
2747627779
- GTIN10603295384861
- Lot number/ GTIN
- Amount recalled
- 9 cases (1/cs)
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
Why it was recalled
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new DePuy Orthopaedics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.