Reported Aug 17, 2016 · Terminated
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
Medline Industries Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medline Industries Inc · 1 Medline Pl · Mundelein, IL · 60060-4485
Is this the product you have?
- Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
- Lot codes and dates
- Lot Numbers 122211502 122212515
- Amount recalled
- 903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
Why it was recalled
There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medline Industries Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.