Reported Jul 31, 2024 · Ongoing
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Beckman Coulter, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter, Inc. · 1000 Lake Hazeltine Dr · Chaska, MN · 55318-1037
Is this the product you have?
Dxl 9000 Access Immunoassay AnalyzerCatalog #C11137
- Lot codes and dates
- Serial numbers
144 of 144 codes
300117300116300137300138300139300129300141300143300142300151300152300128300156300153300154300135300164300180300155300167300166300158300171300179300169300182300176300183300198300192300133300188300161300134300190300209300212300211300149300148300150300130300172300197300160300189300170300181300195300194300224300222300168300210300233300207300205300221300203300240
- Serial numbers
- Amount recalled
- 143 devices
Where it was sold
Why it was recalled
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.