Reported Aug 7, 2024 · Ongoing

SlideView VS200 slide viewer,

Evident Scientific Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: EVIDENT SCIENTIFIC INC · 48 Woerd Ave · Waltham, MA · 02453-3826

Is this the product you have?

  • SlideView VS200 slide viewer,
Lot codes and dates
  • 65159 67105 67104 69153 67106 69154
Amount recalled
N/A

Why it was recalled

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

This product could cause temporary or reversible health problems — stop using it.