Reported Jul 18, 2018 · Terminated

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and…

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 8200 Coral Sea St Ne · Mounds View, MN · 55112-4391

Is this the product you have?

  • Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry
  • D154VRC Product UsageThe implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias
Lot codes and dates

2,708 of 2,708 products

Products covered by this recall
Model / numberProduct as published
T600036S
T600042S
T600053S
T600056S
T600057S
T600059S
T600063S
T600064S
T600067S
T600070S
T600072S
T600077S
T600078S
T600084S
T600086S
T600088S
T600089S
T600090S
T600093S
T600094S
Amount recalled
8821 units

This recall covers 10 products

The agency filed one notice per product. You are reading one of them; the other 9 are below.

Where it was sold

  • Nationwide

Why it was recalled

Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic Inc., Cardiac Rhythm and Heart… recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 16, 2021 · Data as of Sep 19, 5:29 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.