Reported Oct 3, 2012 · Terminated

BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to…

OPTI Medical Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: OPTI Medical Systems, Inc · 235 Hembree Park Dr · Roswell, GA · 30076-5738

Is this the product you have?

  • BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to be used for in vitro measurements of pH, PCOX2 PO2, lactate (lactic acid), total hemoglobin (tHb), and oxygen saturation (SO2), in heparinized whole blood samples (either arterial or venous) on the OPTI CCA-TS system, in either a clinical setting or point of care locations. Cassettes contain one-time use sensors that are used with the OPTI CCA-TS analyzer to perform in-vitro measurements of blood gases and metabolites. OPTI CCA-TS B-Lac Cassettes are supplied in boxes of 25 and may only used with the OPTI CCA-TS Analyzer.
Lot codes and dates
  • Lot numbers
    • 110650
    • 114650
    • 114651
    • 114652
    • 114653
    • 116650
    • 116652
    • 117650
    • 118651
    • 121650
    • 121651
    • 122652
    • 122652
    • 122653
    • 122654
    • 122655
    • 122656
    • 122657
    • 123654
    • 123655
    • 124650
    • 133650
    • 134651
    • 136650
    • 146650
    • 149650
    • 202650
    • 208650
    • 208651
    • 209560
    • 209651
    • 211650
    • 212651
    • 213650
    • 214650
    • 215650
    • 215651
    • 221650
    • 222650 and 224650
Amount recalled
895 boxes (25 cassettes per box)

Where it was sold

Why it was recalled

The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new OPTI Medical Systems, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 11, 2013 · Data as of Sep 19, 5:25 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.