Reported Sep 11, 2019 · Terminated
AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012
Angiodynamics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Angiodynamics, Inc. · 603 Queensbury Ave · Queensbury, NY · 12804-7619
Is this the product you have?
- AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012
- Lot codes and dates
Lot 5456490UPN H787114030120
- Amount recalled
- 109
Where it was sold
Why it was recalled
The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.
This product could cause temporary or reversible health problems — stop using it.