Reported Oct 3, 2012 · Terminated
This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and…
Alcon Research, Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Alcon Research, LTD. · 9965 Buffalo Speedway · Houston, TX · 77054-1309
Is this the product you have?
- This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries.
- Lot codes and dates
- 1) 23G
- TotalPLUS PakAccurus Standalone - 8065750828, 2) 23G Combined Pack 0.9mm: Accurus Standalone - 8065750814, 3) 23G Combined Pack 1.1mm: Accurus Standalone - 8065750816, 4)23G
- TotalPLUS PakConstellation Standalone - 8065751058, 5) 23G Combined Pack 0.9mm: Constellation Standalone -8065751077, 6) 23G Combined Pack 1.1mm: Constellation Standalone - 8065751065. 7) 23G Small Parts Kit - 8065750842. Complete list of lot numbers will be provided with recall recommendation
- Amount recalled
- 93,319 (30, 413 distributed domestically & 62,906 internationally)
Where it was sold
- Nationwide
Why it was recalled
This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Alcon Research, Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.