Reported Sep 6, 2023 · Ongoing

GS70H Salus Surgical Table, Product Code 18-071-70

Skytron, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Skytron, LLC · 5085 Corporate Exchange Blvd SE · Grand Rapids, MI · 49512-5515

Is this the product you have?

    • GS70H Salus Surgical Table
    • Product Code 18-071-70
Lot codes and dates
  • Serial Numbers
    • SUC-20207-0002
    • SUC-20208-0006
    • SUC-20209-0010
    • SUC-2020X-0014
    • SUC-2022Z-0018
    • SUC-20233-0022
    • SUC-20207-0003
    • SUC-20208-0007
    • SUC-20209-0011
    • SUC-2020X-0015
    • SUC-2022Z-0019
    • SUC-20233-0023
    • SUC-20208-0004
    • SUC-20209-0008
    • SUC-2020X-0012
    • SUC-2020X-0016
    • SUC-2022Z-0020
    • SUC-20233-0024
    • SUC-20208-0005
    • SUC-20209-0009
    • SUC-2020X-0013
    • SUC-2022Z-0017
    • SUC-2022Z-0021
    • SUC-20233-0025
Amount recalled
24 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Skytron, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:56 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.