Reported Sep 11, 2019 · Terminated

CE18TKN ACCU-BLOC PERIFIX KIT NRFIT Catalog # 339188

B. Braun Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: B. Braun Medical, Inc. · 901 Marcon Blvd · Allentown, PA · 18109-9512

Is this the product you have?

  • CE18TKN ACCU-BLOC PERIFIX KIT NRFIT Catalog # 339188
Lot codes and dates
  • Lot # 0061537846
  • Exp8/31/2018 Lot # 0061537847
  • Exp10/31/2018 Lot # 0061620282
  • Exp9/30/2019
  • UDI04046964941436
Amount recalled
3820 units

This recall covers 9 products

The agency filed one notice per product. You are reading one of them; the other 8 are below.

Where it was sold

Why it was recalled

The filter capacity is not achieved due to damage of the filter membrane.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new B. Braun Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 21, 2020 · Data as of Sep 19, 5:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.