Reported Sep 22, 2021 · Terminated

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image…

Philips North America LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289

Is this the product you have?

  • The Azurion series (within the limits of the used Operation Room
  • table) are intended for use to perform"Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the
  • following clinical application areasvascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures
  • Additionally"The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow
Lot codes and dates
  • Serial number921
Amount recalled
1 unit

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 11, 2024 · Data as of Sep 19, 7:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.