Reported Sep 6, 2023 · Completed
Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
Accuray Incorporated
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Accuray Incorporated · 1240 Deming Way · Madison, WI · 53717-1954
Is this the product you have?
- Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
- Lot codes and dates
- UDI-DI number 00811376030009. Serial numbers 4010505, 4010506, 4010507, 4010508, 4010509, 4010510, 4010511, 4010513, 4010515, 4010516, 4010517, 4010518, 4010519, 4010521, 4010522, 4010523, 4010524, 4010525, 4010526, 4010527, 4010528, 4010529, 4010530, 4010531, 4010532, 4010533, 4010536, 4010537, 4010538, 4010539, 4010541, 4010542, 4010543, 4010544, 4010545, 4010546, 4010548, 4010549, 4010550, 4010551, 4010552, 4010553, 4010554, 4010555, 4010556, 4010557, 4010558, 4010559, 4010560, 4010561, 4010562, 4010563, 4010564, 4010565, 4010566, 4010567, 4010568, 4010569, 4010570, 4010571, 4010573, 4010574, 4010577, 4010579, 4010582, 4010583, 4010584, 4010585, 4010586, 4010587, 4010588, 4010589, 4010590, 4010592, 4010593, 4010594, 4010596, 4010599, 4010607, 4010600, 4010601, and 4010609.
- Amount recalled
- 82
Where it was sold
Why it was recalled
It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Accuray Incorporated recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.