Reported Jul 15, 2020 · Ongoing

TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Software Versions…

Tomtec Imaging Systems GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Tomtec Imaging Systems Gmbh · Edisonstr. 6 · Unterschleisheim · N/A

Is this the product you have?

    • TOMTEC-ARENA TTA2
    • software version no. 2.20 and lower
    • Picture archiving and communications system
  • Software Versions and applicable UDITTA 1.2 release prior to UDI requirement TTA2.00.01 release prior to UDI requirement TTA2.11.01 release prior to UDI requirement TTA2.11.02 release prior to UDI requirement TTA2.20.01
  • UDI(01)04260361440143(11)160429(10)20.01 TTA2.20.10
  • UDI(01)04260361440143(11)160809(10)20.10 TTA2.20.13
  • UDI(01)04260361440143(11)160922(10)20.13 TTA2.20.14
  • UDI(01)04260361440143(11)161028(10)20.14 TTA2.20.15
  • UDI(01)04260361440143[11]161207[10]20.15 TTA2.20.16
  • UDI(01)04260361440143(11)170127(10)20.16 TTA2.20.18
  • UDI(01)04260361440143(11)170622(10)20.18
Lot codes and dates
  • The affected product versions are TOMTEC-ARENA, TTA 2.20 and lower, with modules 4D LV-Analysis, 4D RV-Function, 4D MV-Assessment, 3D Cardio-View, 4D Sono-Scan, Echo-Com, 2D Cardiac Performance Analysis, and QAngio.
Amount recalled
62 units

Where it was sold

Why it was recalled

The firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Tomtec Imaging Systems GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.