Reported Aug 1, 2018 · Terminated

UNIQ Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional…

Philips Healthcare

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Healthcare · 3000 Minuteman Rd · Andover, MA · 01810-1032

Is this the product you have?

  • UNIQ Product UsageVascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures
Lot codes and dates
  • UNIQ 1.0.10 UNIQ 1.0.10.5
Amount recalled
459 affected systems

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Healthcare recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 3, 2024 · Data as of Sep 19, 4:47 PM UTC