Reported Aug 14, 2024 · Ongoing

TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with…

Integra LifeSciences Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. · 1100 Campus Rd · Princeton, NJ · 08540-6650

Is this the product you have?

  • TruDi NAV Suction, 0 Degrees
  • Model/Catalog NumberTDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
Lot codes and dates
  • UDI/DI10705031245877
  • Lot Codes
    • 2310160
    • 2309110
    • 2309150
    • 2308140
    • 2308160
    • 2308180
Amount recalled
265 units

Where it was sold

Why it was recalled

Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.