TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with…
Integra LifeSciences Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Integra LifeSciences Corp. · 1100 Campus Rd · Princeton, NJ · 08540-6650
Is this the product you have?
- TruDi NAV Suction, 0 Degrees
- Model/Catalog NumberTDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
- Lot codes and dates
- UDI/DI10705031245877
- Lot Codes
231016023091102309150230814023081602308180
- Amount recalled
- 265 units
Where it was sold
Why it was recalled
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Integra LifeSciences Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.