Reported Sep 9, 2015 · Terminated
KimVent* Turbo-Cleaning Closed Suction System for Adults, 14 F, T-Piece (Product Code 22705); KimVent* Turbo-Cleaning…
Halyard Health, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Halyard Health, Inc · 5405 Windward Pkwy · Alpharetta, GA · 30004-3894
Is this the product you have?
- KimVent* Turbo-Cleaning Closed Suction System for Adults, 14 F, T-Piece (Product Code 22705); KimVent* Turbo-Cleaning Closed Suction System for Adults, 14 F, T-Piece (Product Code 2270135) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
- Lot codes and dates
Product Code 22705 - M5082T301M5096T315Product Code 2270135 - M5076T321M5082T306
- Amount recalled
- N/A
This recall covers 32 products
The agency filed one notice per product. You are reading one of them; the other 31 are below.
19 more are in the archive.
Where it was sold
- Nationwide
- PR
Why it was recalled
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Halyard Health, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.