Reported Sep 10, 2025 · Ongoing

Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL,…

ICU Medical Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ICU Medical Inc. · 6000 Nathan Ln N · Minneapolis, MN · 55442-1690

Is this the product you have?

  • Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7002-24
Lot codes and dates
  • UDI/DI 10610586022128
  • Lot Numbers
    • 6022010
    • 6026827
    • 6026828
    • 6026829
    • 6026830
    • 6026831
    • 6037750
    • 6037751
    • 6037752
    • 6037753
    • 6053980
    • 6053981
    • 6053982
    • 6053983
    • 6062689
    • 6062688
    • 6063845
    • 6062691
Amount recalled
265500 units

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Why it was recalled

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new ICU Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:41 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.