Reported Sep 9, 2015 · Terminated

VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and…

Ortho-Clinical Diagnostics

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ortho-Clinical Diagnostics · 100 Indigo Creek Dr · Rochester, NY · 14626-5101

Is this the product you have?

  • VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644, IVD --- Ortho Clinical Diagnostics Inc. The VITROS 5,1 FS Chemistry System with enGEN Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Lot codes and dates
    • Software Versions 2.2.1 through 2.8
    • Catalog # 6801375
  • Serial Numbers
    • 34000119 through 34002323
    • Catalog # 6801890
  • Serial Numbers

    63 of 63 codes

    • J34000133
    • J34000193
    • J34000197
    • J34000212
    • J34000285
    • J34000316
    • J34000318
    • J34000337
    • J34000360
    • J34000420
    • J34000451
    • J34000478
    • J34000495
    • J34000497
    • J34000529
    • J34000538
    • J34000552
    • J34000557
    • J34000590
    • J34000596
    • J34000601
    • J34000631
    • J34000642
    • J34000761
    • J34000799
    • J34000805
    • J34000833
    • J34000851
    • J34000893
    • J34000946
    • J34000947
    • J34001006
    • J34001172
    • J34001218
    • J34001240
    • J34001243
    • J34001248
    • J34001251
    • J34001274
    • J34001297
    • J34001311
    • J34001317
    • J34001345
    • J34001378
    • J34001379
    • J34001389
    • J34001390
    • J34001396
    • J34001422
    • J34001472
    • J34001504
    • J34001569
    • J34001572
    • J34001612
    • J34001626
    • J34001633
    • J34001668
    • J34001755
    • J34001809
    • J34001840
Amount recalled
Catalog 6801375: Domestic 919; Foreign 1251; Catalog 6801890: Domestic 63, Foreign 147

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2.1 through 2.8. If the user is utilizing User- Configured (Modified) parameters, ALL default settings are restored in the VITROS 5,1 FS Chemistry System from the ADD instead of retaining the user modified parameters.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 29, 2018 · Data as of Sep 19, 6:06 PM UTC