Reported Aug 1, 2018 · Terminated

RayStation Radiation Therapy Treatment Planning System, Model nos. 2.5, 3.5, 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1,…

Raysearch Laboratories Ab

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: RAYSEARCH LABORATORIES AB · Sveavagen 9 · Stockholm, N/A · N/A

Is this the product you have?

  • RayStation Radiation Therapy Treatment Planning System, Model nos. 2.5, 3.5, 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 7.0 Product Usage RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Lot codes and dates
    • 2.5.1.89
    • 3.5.0.16
    • 3.5.1.6
    • 4.0.0.14
    • 4.0.1.4 4.0.2.9
    • 4.0.3.4
    • 4.3.0.14
    • 4.5.0.19
    • 4.5.1.14
    • 4.5.2.7
    • 4.7.0.15
    • 4.7.1.10
    • 4.7.2.5
    • 4.7.3.13
    • 4.7.4.4
    • 4.7.5.4
    • 4.9.0.42
    • 5.0.0.37
    • 5.0.1.11
    • 5.0.2.35
    • 6.0.0.24
    • 6.1.0.26
    • 6.1.1.2
    • 6.2.0.7
    • 7.0.0.19
Amount recalled
718

Where it was sold

  • Nationwide

Why it was recalled

The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Raysearch Laboratories Ab recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 4, 2019 · Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.