Reported Oct 6, 2021 · Terminated

DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634,…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

    • DxA 5000 (DxA Automation System
    • DxA 5000 fit)
    • "For in Vitro Diagnostic Use"
  • Reference NumbersB87352 with B57634
  • UDI15099590707118
Lot codes and dates
  • Affected US Serial Numbers

    76 of 76 codes

    • T23000077
    • T23000075
    • T23000076
    • T23000074
    • T23000069
    • T23000070
    • T23000066
    • T23000099
    • T23000096
    • T23000098
    • T23000100
    • T23000095
    • T23000097
    • T23000101
    • T23000092
    • T23000091
    • T23000090
    • T23000142
    • T23000045
    • T23000040
    • T23000039
    • T23000038
    • T23000152
    • T23000154
    • T23000155
    • T23000153
    • T23000106
    • T23000108
    • T23000107
    • T23000109
    • T23000128
    • T23000124
    • T23000125
    • T23000126
    • T23000115
    • T23000116
    • T23000113
    • T23000112
    • T23000117
    • T23000114
    • T23000132
    • T23000133
    • T23000131
    • T23000134
    • T23000064
    • T23000065
    • T23000063
    • T23000062
    • T23000163
    • T23000164
    • T23000162
    • T23000173
    • T23000169
    • T23000161
    • T23000138
    • T23000136
    • T23000137
    • T23000094
    • T23000093
    • T23000135
Amount recalled
123 systems

Where it was sold

Why it was recalled

There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 10, 2024 · Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.