Reported Jul 8, 2026 · Ongoing
Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F…
Medline Industries, LP
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medline Industries, LP · 3 Lakes Dr · Northfield, IL · 60093-2753
Is this the product you have?
- Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
- Lot codes and dates
- DYNJ905503D UDI-DI 10193489430295 lots 21DBG611 21EBI605 21FBN751 DYNJ905503F UDI-DI 10193489967272 lotsUDI-DI 10198459200397 lots
21HBL06021IBW85922DBC34022DBS43722FBR26722GBR28622HBS56022IBA57722JBI88322KBD58722LBF42523BBN90323CBL45023CBW77623FBC20923IBU34324DBQ18124FBT06224GBR74024HBI42824JBP984DYNJ905503G
25ABJ06925ABL74925CBB76225DBG79925EBM99425GBW22625LBA95025LBT01926ABH51926ABT31526BBQ33826DBJ110
- DYNJ905503D UDI-DI 10193489430295 lots 21DBG611 21EBI605 21FBN751 DYNJ905503F UDI-DI 10193489967272 lots
- Amount recalled
- 9,272 kits
This recall covers 24 products
The agency filed one notice per product. You are reading one of them; the other 23 are below.
11 more are in the archive.
Where it was sold
Why it was recalled
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medline Industries, LP recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.