Reported Sep 17, 2014 · Terminated
The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of…
Becton Dickinson & Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson & Co. · BD Diagnostic Systems · Sparks, MD · 21152-0999
Is this the product you have?
- The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.
- Lot codes and dates
All Lots with the expiration date January 212014 or later
- Amount recalled
- 585,048 tests
Where it was sold
Why it was recalled
The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new produce recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.