Reported Oct 1, 2025 · Ongoing

Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638,…

Philips Medical Systems Nederland B.v.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 4-6 · Best, N/A · N/A

Is this the product you have?

  • Allura Xper FD10 OR Table Product Codes
    • (1)722022
    • (2) 722033
  • Associated 510(k) numbers
    • K102005
    • K130842
    • K130638
    • K133292
    • K141979
    • K161563
    • K162859
    • K033737
    • K041949
Lot codes and dates
  • Product Codes
    • (1)722022
    • (2) 722033
  • UDI-DI
    • (1) 00884838059078
    • (2) 00884838054233
  • Serial numbers
    • (1) 6
    • (2) 1
Amount recalled
2 units (1 US, 1 OUS)

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Where it was sold

  • Nationwide
  • GA

Why it was recalled

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Medical Systems Nederland B.v. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:58 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.