Reported Oct 4, 2023 · Ongoing
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air…
3M Company - Health Care Business
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: 3M Company - Health Care Business · 3m Center 2510 Conway Ave · Saint Paul, MN · 55144-0001
Is this the product you have?
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
- Lot codes and dates
- UDI/DI50707387787283( (shipper)
- Lot Codes
41482VFexp. 33P4ND33P4YNexp. 01/02/2025
- Amount recalled
- 77,280 devices
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new 3M Company - Health Care Business recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.