Reported Oct 4, 2023 · Ongoing

BD Pyxis MedBank MedPass Software, REF: 139088-01

CareFusion 303, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: CareFusion 303, Inc. · 10020 Pacific Mesa Blvd · San Diego, CA · 92121-4386

Is this the product you have?

  • BD Pyxis MedBank MedPass Software
  • REF139088-01
Lot codes and dates
  • UDI-DI10885403513817, Software versions prior to but not including v1.2.3.23
Amount recalled
37

Where it was sold

Why it was recalled

After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new CareFusion 303, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:39 PM UTC