Reported Jul 5, 2017 · Terminated

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China…

Carestream Health Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Carestream Health Inc · 150 Verona St · Rochester, NY · 14608-1733

Is this the product you have?

3 of 3 products

Products covered by this recall
Model / numberProduct as published
DRX-Revolution Mobile X-Ray System
DRXR-1
1060177Catalog # 1019397 (all other countries) and (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures
Lot codes and dates
  • Serial NumbersWorldwide
  • except China101 through 2623
  • China only800101 through 800482
Amount recalled
Domestic: 1449 units; Foreign: 1416 units

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Carestream Health Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 13, 2018 · Data as of Sep 19, 5:34 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.