Reported Sep 24, 2014 · Terminated

4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery…

Varian Medical Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Varian Medical Systems, Inc. · 911 Hansen Way · Palo Alto, CA · 94304-1028

Is this the product you have?

  • 4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery by monitoring linear accelerator parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan.
Lot codes and dates
  • Model number H51
  • Serial numbers

    512 of 512 codes

    • H51B429
    • H514517
    • H511720
    • H514463
    • H511002
    • H513925
    • H514115
    • H510855
    • H515068
    • H514544
    • H512492
    • H515879
    • H515060
    • H513064
    • H514438
    • H514022
    • H512957
    • H514553
    • H515102
    • H515991
    • H515511
    • H51B421
    • H515106
    • H514189
    • H513551
    • H513009
    • H515033
    • H51B568
    • H515804
    • H51B611
    • H515004
    • H513477
    • H513133
    • H513562
    • H510966
    • H515691
    • H513894
    • H515363
    • H513801
    • H51B362
    • H518326
    • H518191
    • H514863
    • H513889
    • H514050
    • H515619
    • H515282
    • H515413
    • H51B240
    • H515815
    • H51B204
    • H512299
    • H513438
    • H51B233
    • H510451
    • H512146
    • H512507
    • H515601
    • H513692
    • H51B136
Amount recalled
512 affected devices

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • GA

Why it was recalled

Due to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian's Bar Code Conical Verification system from the BrainLab's ExacTract console when it is used with BCCV.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 30, 2015 · Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.