Reported Jul 12, 2017 · Terminated
Alaris PC Unit, Model 8015
CareFusion 303, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: CareFusion 303, Inc. · 10020 Pacific Mesa Blvd · San Diego, CA · 92121-4386
Is this the product you have?
- Alaris PC Unit, Model 8015
- Lot codes and dates
- Notification will be distributed to all Alaris PC Unit model 8015 customers with software versions greater than 9.12.
- Amount recalled
- 568,283 units
Where it was sold
- Nationwide
Why it was recalled
BD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new CareFusion 303, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.