Reported Jul 29, 2020 · Terminated
Ablation Galil Technology, Three IceFORCE 2.1 CX Prostate Cyroablation Kit Visual ICE System, Prescription Only Galil…
Boston Scientific Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Boston Scientific Corporation · 300 Boston Scientific Way · Marlborough, MA · 01752-1291
Is this the product you have?
- Ablation Galil Technology, Three IceFORCE 2.1 CX Prostate Cyroablation Kit Visual ICE System, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.
- Lot codes and dates
- UPNFPRPR3607
- GTIN7290104831058
- Batch
A1194A1298A1436A1569A1567A1782A6754A6757A6848A6849A6915T0154T0163T0552U0083
- Amount recalled
- 88 devices
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
Why it was recalled
The needle surface deteriorated due to unexpected electrolysis during procedure.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Boston Scientific Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.