Reported Sep 7, 2016 · Terminated

Monaco RTP System Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Elekta, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Elekta, Inc. · 400 Perimeter Center Ter NE Ste 50 · Atlanta, GA · 30346-1227

Is this the product you have?

  • Monaco RTP System Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Lot codes and dates
  • Software 3.1 and higher.
Amount recalled
3,012 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

In a specific workflow where contours are edited (enlarged, moved, copied, etc.) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Elekta, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 28, 2025 · Data as of Sep 19, 5:20 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.