Reported Sep 14, 2016 · Terminated

ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

ARJOHUNTLEIGH POLSKA Sp. z.o.o.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ARJOHUNTLEIGH POLSKA Sp. z.o.o. · UI. KS. PIOTRA WAWRZYNIAKA 2 · KOMORNIKI · N/A

Is this the product you have?

  • ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.
Lot codes and dates
  • 108 of 108 codes

    • Device Listing Number E101987
    • Model Number APA1000-XX
    • Serial numbers P0359671
    • P0360103
    • P0338345
    • P0354548
    • P0357690
    • P0359672
    • P0359673
    • P0348548
    • P0349629
    • P0361336
    • P0361337
    • P0361875
    • P0361876
    • P0361877
    • P0360104
    • P0360105
    • P0346000
    • P0373592
    • P0357691
    • P0349634
    • P0354549
    • P0349627
    • P0339257
    • P0321458
    • P0299662
    • P0325930
    • P0325931
    • P0325933
    • P0310854
    • P0344175
    • P0356357
    • P0356362
    • P0356363
    • P0360990
    • P0375493
    • P0298175
    • P0309635
    • P0310848
    • P0323622
    • P0323624
    • P0325167
    • 1404000118
    • P0296131
    • P0296315
    • P0298465
    • P0319583
    • P0333285
    • P0333288
    • P0339264
    • P0343583
    • P0343584
    • P0343585
    • P0343684
    • P0344174
    • P0353246
    • P0364628
    • 1404000112
    • P0333284
Amount recalled
106

Where it was sold

Why it was recalled

ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees¿-28 degrees¿. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees¿ and the device is not loaded over 330 lbs (150 kg).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new ARJOHUNTLEIGH POLSKA Sp. z.o.o. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 8, 2019 · Data as of Sep 19, 6:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.