Reported Oct 8, 2014 · Terminated
Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior…
Instrumed International, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Instrumed International, Inc. · 626 Cooper Ct · Schaumburg, IL · 60173-4537
Is this the product you have?
- Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
- Lot codes and dates
4 of 4 products
Products covered by this recall Model / number Product as published 250-08860Instrumed ; Symmetry Surgical Inc. Catalog Code: 90-9525; Lot Number: 030214 250-08930Instrumed ; Symmetry Surgical Inc. Catalog Code: 90-9522; Lot Number: 070511 250-08990Instrumed ; Symmetry Surgical Inc. Catalog Code: 90-9526; Lot Number: 031511 250-09010Instrumed ; Symmetry Surgical Inc. Catalog Code: 90-9523; Lot Number: 080112 - Amount recalled
- 895 instruments
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
Why it was recalled
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Instrumed International, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.