Reported Sep 4, 2024 · Ongoing
DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Beckman Coulter, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter, Inc. · 1000 Lake Hazeltine Dr · Chaska, MN · 55318-1037
Is this the product you have?
DxI 9000 Access Immunoassay AnalyzerCatalog #C11137
- Lot codes and dates
Software versions 1.17.0 or belowUDI-DI 15099590732103
- Amount recalled
- 183 analyzers
Where it was sold
Why it was recalled
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.