Reported Sep 4, 2024 · Ongoing
Afinion 2 Analyzer, model #1116986, US refurbished, CLIA waived.
Abbott Diagnostics Technologies AS
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Diagnostics Technologies AS · Kjelsasveien 161 · Oslo · N/A
Is this the product you have?
- Afinion 2 Analyzer, model #1116986, US refurbished, CLIA waived.
- Lot codes and dates
Lot #433706Serial numbers AF20008880AF20002691AF20035414AF20015543 and AF20035460
- UDI-DI07070060014708
- Amount recalled
- 5 analyzers
Where it was sold
Why it was recalled
Five Afinion 2 (AF2) Analyzer were incorrectly market configured as Moderate Complexity and distributed to customers. Use of moderately complex assays by waived laboratories could result in erroneous results.