Reported Sep 16, 2015 · Terminated
CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial…
Terumo Cardiovascular Systems Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Terumo Cardiovascular Systems Corporation · 6200 Jackson Rd · Ann Arbor, MI · 48103-9586
Is this the product you have?
- CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature.
- Lot codes and dates
- 500A, 500AHCT, 500AV, 500AVHCT CDI Blood Parameter Monitoring System 500; software version 1.69.
- Amount recalled
- 4638
Where it was sold
Why it was recalled
Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Terumo Cardiovascular Systems Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.