Reported Jul 22, 2026 · Ongoing

Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Daavlin Distributing Company

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Daavlin Distributing Company · 205 W Bement St · Bryan, OH · 43506-1264

Is this the product you have?

  • Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Lot codes and dates
  • Serial Numbers

    212 of 212 codes

    • 6-09278NB8CL
    • 6-09271NB8CL
    • 6-09055NB8CL
    • 6-09113NB8CL
    • 6-09204NB8CL
    • 6-09207NB8CL
    • 6-09116NB8CL
    • 6-09263NB8CL
    • 6-09272NB8CL
    • 6-09118NB8CL
    • 6-09275NB8CL
    • 6-09264NB8CL
    • 6-09060NB8CL
    • 6-09122NB8CL
    • 6-09216NB8CL
    • 6-09125NB8CL
    • 6-09353NB8CL
    • 6-09346NB8CL
    • 6-09128NB8CL
    • 6-09129NB8CL
    • 6-09133NB8CL
    • 6-09134NB8CL
    • 6-09266NB8CL
    • 6-09215NB8CL
    • 6-09355NB8CL
    • 6-09137NB8CL
    • 6-09359NB8CL
    • 6-09343NB8CL
    • 6-09140NB8CL
    • 6-09142NB8CL
    • 6-09143NB8CL
    • 6-09273NB8CL
    • 6-09065NB8CL
    • 6-09144NB8CL
    • 6-09237NB8CL
    • 6-09225NB8CL
    • 6-09145NB8CL
    • 6-09147NB8CL
    • 6-09148NB8CL
    • 6-09378NB8CL
    • 6-09252NB8CL
    • 6-09051NB8CL
    • 6-09260NB8CL
    • 6-09250NB8CL
    • 6-09149NB8CL
    • 6-09238NB8CL
    • 6-09212NB8CL
    • 6-09277NB8CL
    • 6-09048NB8CL
    • 6-09274NB8CL
    • 6-09259NB8CL
    • 6-09224NB8CL
    • 6-09057NB8CL
    • 6-09436NB12CL
    • 6-09438NB8CL
    • 6-09434NB12CL
    • 6-09270NB8CL
    • 6-09115NB8CL
    • 6-09111NB8CL
    • 6-09435NB12CL
Amount recalled
212 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Daavlin Distributing Company recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.