Reported Jul 26, 2017 · Terminated

BD 5ml Syringe Luer-Lok" Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #305062

Becton Dickinson & Company

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Becton Dickinson & Company · 1 Becton Dr · Franklin Lakes, NJ · 07417-1815

Is this the product you have?

  • BD 5ml Syringe Luer-Lok" Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #305062
Lot codes and dates
  • 6204878 2021-07
Amount recalled
22,400 units

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide

Why it was recalled

BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Becton Dickinson & Company recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 20, 2018 · Data as of Sep 19, 4:46 PM UTC