Reported Sep 23, 2015 · Terminated

Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one…

Pointe Scientific, Inc.

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Pointe Scientific, Inc. · 5449 Research Dr · Canton, MI · 48188-2261

Is this the product you have?

  • Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle
  • R1HDPE plastic bottle. 120 ml fill
  • R2HDPE plastic bottle. 90 ml fill
Lot codes and dates
  • C7539-150 Lot406201
  • Batch279
  • Expiration date2016-03
Amount recalled
93 kits

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Pointe Scientific, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 29, 2015 · Data as of Sep 19, 6:16 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.