Reported Sep 23, 2015 · Terminated
Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and…
Fujifilm Medical Systems U.S.A., Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Fujifilm Medical Systems U.S.A., Inc. · 10 Highpoint Dr · Wayne, NJ · 07470-7431
Is this the product you have?
- Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.
- Lot codes and dates
- ED-530XT - Part #202B1237697D; ED-250XT5 and ED-250XL5 - Part #202B1260196A; ED-450XT5 - Part #202B1260211A and ED-450XL5 - Part #202B1237681
- Amount recalled
- 256 units
Why it was recalled
The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.