Reported Sep 23, 2015 · Terminated
The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue…
Leica Microsystems, Inc.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Leica Microsystems, Inc. · 1700 Leider Ln · Buffalo Grove, IL · 60089-6622
Is this the product you have?
- The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
- Lot codes and dates
- Product CodeLeica
- ASP200S/ASP300S with Serial Numbers
79 of 79 codes
568258865264526753565259595659465794538559295301540153105530590958045272572457965807544654355670594753485434538853666000613154076040579958155737570653815416567358755962604160425266594856936021561057985926538753495273611556125965595552605725
- Amount recalled
- 901
Where it was sold
Why it was recalled
The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Leica Microsystems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.