Reported Sep 26, 2018 · Terminated

Temperature sensing catheter 14FR, Catalogue Number 102201101463NU Product Usage: This is sterile, single use, two-way…

Degania Silicone, Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Degania Silicone, Ltd. · Degania Bet · Dganya Bet, N/A · N/A

Is this the product you have?

  • Temperature sensing catheter 14FR, Catalogue
  • Number 102201101463NU Product UsageThis is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period
Lot codes and dates
  • V17012949
Amount recalled
1000

This recall covers 32 products

The agency filed one notice per product. You are reading one of them; the other 31 are below.

19 more are in the archive.

Where it was sold

Why it was recalled

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Degania Silicone, Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 25, 2024 · Data as of Sep 19, 5:31 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.