Reported Aug 22, 2018 · Terminated
MegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715
Megadyne Medical Products, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Megadyne Medical Products, Inc. · 11506 S State St · Draper, UT · 84020-9453
Is this the product you have?
MegaPower Electrosurgical GeneratorProduct code 1000GTIN 10614559103715
- Lot codes and dates
3,279 of 3,279 codes
000010010001000113806001000530010006300100066001000700010007800100085001000910010010000100105001001100010011500100128001001320010014600100167001001810010018700100191001002000010021100100212001002230010022900100237001002400010024200100246001002480010025400100265001002780010028200100283001003080010031300100313001003140010033300100334001003790010038000100391001003980010040200100413001004160010041800100424001004320010043700100438001004580010045900100471001004790010048100100484001
- Amount recalled
- 8380 units
Where it was sold
- Nationwide
Why it was recalled
If two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Megadyne Medical Products, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.